Try the modernized ClinicalTrials.gov beta website. Learn more about the modernization effort.
Working…
ClinicalTrials.gov
ClinicalTrials.gov Menu

M5 vs. M6 Comparison Study With a Sub Study Into the Dielectric Constant of Aspirated Cyst Fluid

The safety and scientific validity of this study is the responsibility of the study sponsor and investigators. Listing a study does not mean it has been evaluated by the U.S. Federal Government. Read our disclaimer for details.
 
ClinicalTrials.gov Identifier: NCT03771833
Recruitment Status : Completed
First Posted : December 11, 2018
Last Update Posted : February 20, 2020
Sponsor:
Information provided by (Responsible Party):
Micrima, Ltd.

Brief Summary:
The MARIA breast imaging system is a Conformité Européenne (CE)-marked radio-frequency (RF) medical imaging device. The system employs an electromagnetic imaging technique that exploits the dielectric contrast between normal and cancerous tissues. The performance and imaging characteristic differences between the M5 and M6 versions of MARIA are not yet well demonstrated in the clinical environment, particularly with regards to cysts. The evaluation of some aspects of this potentially important new technology will occur in this comparative technical study. Further, the dielectric constant of cyst fluid is currently not well understood and obtaining readings from aspirated cyst fluid in applicable patients will be attempted.

Condition or disease Intervention/treatment Phase
Breast Cyst Breast Cancer Female Device: MARIA scan Not Applicable

Layout table for study information
Study Type : Interventional  (Clinical Trial)
Actual Enrollment : 22 participants
Allocation: Non-Randomized
Intervention Model: Parallel Assignment
Masking: None (Open Label)
Primary Purpose: Other
Official Title: Comparison of M5 and M6 Versions of the MARIA Imaging System on Patients Attending Symptomatic Breast Clinic in Cheltenham, United Kingdom (UK), Including a Sub-study to Research the Dielectric Constant of Aspirated Cyst Fluid
Actual Study Start Date : February 19, 2019
Actual Primary Completion Date : August 19, 2019
Actual Study Completion Date : January 23, 2020

Resource links provided by the National Library of Medicine

MedlinePlus related topics: Nuclear Scans

Arm Intervention/treatment
Main MARIA scan visit
For Arm 1, participants will be identified in clinic as having a suitable symptomatic breast and approached about the study. They will have the study design explained to them and will be informed that they are under no obligation to participate. Arm 1 participants will receive study information that they will be sent home with and informed that they will be approached via a telephone call in 2-3 days (or the closest working day to that date) to enquire if they would like to schedule an appointment for the study visit. If so, this will be scheduled to occur around 7 days from the date of the phone call.
Device: MARIA scan
The MARIA breast imaging system comprises of a scanning unit, housed under a patient bed. The scanning unit houses a hemispherical array, into which varying sizes of insert are fitted to ensure a close fit to the breast. The patient lies prone on the bed with their breast pendulous through an aperture in the bed and the scanning unit is risen to meet the breast. The fit is non-compressing and the scanner does not use ionising radiation.

Same-day MARIA scan visit

For Arm 2, participants will be identified in clinic as having a suitable symptomatic breast and approached about the study. They will have the study design explained to them and will be informed that they are under no obligation to participate. Arm 2 participants will receive study information and as much time as possible to consider their involvement with the study (at least 1 hour). If the patient agrees to participate, they will be scheduled to have their MARIA scans at a time that suits their commitments that day.

This arm also includes the 2b group, who can optionally consent to a dielectric constant reading of their routinely-aspirated cyst fluid before this is disposed of as per usual site process.

Device: MARIA scan
The MARIA breast imaging system comprises of a scanning unit, housed under a patient bed. The scanning unit houses a hemispherical array, into which varying sizes of insert are fitted to ensure a close fit to the breast. The patient lies prone on the bed with their breast pendulous through an aperture in the bed and the scanning unit is risen to meet the breast. The fit is non-compressing and the scanner does not use ionising radiation.




Primary Outcome Measures :
  1. Establish the presence or absence of a difference in the MARIA(R) outputs obtained with M5 and M6 when the participants is not moved between scans and when they are moved between scans [ Time Frame: Up to 12 weeks ]
    This will be established from Arm 1 and Arm 2 by scanning the participant between both versions of the MARIA system. As has been investigated in-house with healthy volunteers, the patient will take part in a series of scans that will involve them remaining stationary between scans, and moving between scans. The differences between these images will be analysed, using the known palpable lesion in the breast as a landmark.

  2. Establish the presence or absence of a difference between MARIA(R) outputs on the same device when the participant is stationary between scans and when they are moved between scans [ Time Frame: Up to 12 weeks ]
    This will be established from Arm 1 and Arm 2 by scanning the participant on each separate version of the MARIA system. As has been investigated in-house with healthy volunteers, the patient will take part in a series of scans that will involve them remaining stationary between scans, and moving between scans. The differences between these images will be analysed, using the known palpable lesion in the breast as a landmark.

  3. Measure the dielectric constant of aspirated cyst fluid and any variance between cyst type [ Time Frame: Up to 12 weeks ]
    This will be established from Arm 2b by recording the dielectric constant of the aspirated cyst fluid as soon as possible after aspiration using a dielectric probe. No cysts will be aspirated exclusively for this procedure and fluid will be disposed of as per usual site practice once the recording has taken place. Each cyst fluid sample will have its dielectric constant recorded three times and recorded in the provided software before disposal.


Secondary Outcome Measures :
  1. Assess the diagnostic accuracy of both the M5 and M6 versions of the MARIA system in identifying lesions of the breast, including benign lesions such as cysts [ Time Frame: Up to 12 weeks ]
    This will be established from Arm 1 and 2 by using a combination of reader results from clinician reads performed by the site team with all available clinical information and blind reads performed in house which will utilise no clinical information.

  2. Understand participant acceptability of the MARIA scan and obtain feedback on patient experience: questionnaire [ Time Frame: Up to 12 weeks ]
    This will be established from Arm 1 and 2 by asking the participant to complete a questionnaire of their experience after their participation in either of the two main study arms. The questionnaire comprises of 8 questions and a free-text box for any additional comments. The first two questions ask a participant to pick one of 5 options, from very easy to very difficult or very comfortable to very uncomfortable. The next even questions are Yes or No questions, with options for the participant to answer these as "Not sure" or "Not applicable" as appropriate. The questionnaire then ends with a free-text box for feedback.



Information from the National Library of Medicine

Choosing to participate in a study is an important personal decision. Talk with your doctor and family members or friends about deciding to join a study. To learn more about this study, you or your doctor may contact the study research staff using the contacts provided below. For general information, Learn About Clinical Studies.


Layout table for eligibility information
Ages Eligible for Study:   18 Years to 80 Years   (Adult, Older Adult)
Sexes Eligible for Study:   Female
Gender Based Eligibility:   Yes
Gender Eligibility Description:   The device is currently registered for the scanning of female human breast tissue only
Accepts Healthy Volunteers:   No
Criteria

Inclusion Criteria:

  • attending a symptomatic breast care clinic at the study site
  • have a palpable lump in the breast
  • female sex
  • 18 years or older
  • able to provide informed consent
  • not in any identified vulnerable group

Exclusion Criteria:

  • unable to mount MARIA patient bed using provided 2-step
  • unable to lie in the prone position for a period of up to 15 minutes
  • patient who have undergone biopsy less than 5 days before the MARIA scan
  • patient with implanted electronics
  • patient with breast implants
  • patients with nipple piercings (unless they are removed prior to the MARIA scan)
  • breast size too small or too large to be suitable for both arrays

Further:

  • participants with a large cyst or cysts that require draining at their clinical appointment will not be eligible for Arm 1, as the procedure to drain their cyst will change the appearance of the cyst from the imaging that they have done that day to the MARIA image obtained around 10 days later
  • Arm 1 and 2 are mutually exclusive

Information from the National Library of Medicine

To learn more about this study, you or your doctor may contact the study research staff using the contact information provided by the sponsor.

Please refer to this study by its ClinicalTrials.gov identifier (NCT number): NCT03771833


Locations
Layout table for location information
United Kingdom
Thirlestaine Breast Centre, Gloucestershire Hospitals NHS Foundation Trust
Cheltenham, United Kingdom, GL53 7AS
Sponsors and Collaborators
Micrima, Ltd.
Layout table for additonal information
Responsible Party: Micrima, Ltd.
ClinicalTrials.gov Identifier: NCT03771833    
Other Study ID Numbers: PROT-P2-024
First Posted: December 11, 2018    Key Record Dates
Last Update Posted: February 20, 2020
Last Verified: February 2020
Individual Participant Data (IPD) Sharing Statement:
Plan to Share IPD: No
Plan Description: Data collected will be anonymised

Layout table for additional information
Studies a U.S. FDA-regulated Drug Product: No
Studies a U.S. FDA-regulated Device Product: No
Keywords provided by Micrima, Ltd.:
breast
cancer
breast cancer
imaging
Additional relevant MeSH terms:
Layout table for MeSH terms
Cysts
Breast Cyst
Neoplasms
Breast Diseases
Skin Diseases
Pathological Conditions, Anatomical