Nonin 4 Wavelength Cerebral Oximeter Study (Nonin4)
|
The safety and scientific validity of this study is the responsibility of the study sponsor and investigators. Listing a study does not mean it has been evaluated by the U.S. Federal Government. Read our disclaimer for details. |
| ClinicalTrials.gov Identifier: NCT01762722 |
|
Recruitment Status :
Completed
First Posted : January 8, 2013
Results First Posted : April 16, 2013
Last Update Posted : April 16, 2013
|
- Study Details
- Tabular View
- Study Results
- Disclaimer
- How to Read a Study Record
| Condition or disease | Intervention/treatment | Phase |
|---|---|---|
| Healthy | Device: Reduction of inspired oxygen | Not Applicable |
This is a calibration and validation study of a near-infrared spectroscopy (NIRS) device designed to measure the cerebral tissue oxygen saturation non-invasively. This is achieved by comparing NIRS-derived cerebral tissue oxygen saturation with a calculated value derived from simultaneous arterial and jugular venous blood samples.
At present the FDA have adopted the standards published in 2005 by the International Organization for Standardization (ISO), entitled ISO 9919. This is a set of technical specifications and guidelines for pulse oximeters, which share certain technical similarities to cerebral oximeters. In particular, Annex EE details the conduct of a controlled desaturation study for the calibration of pulse oximeter equipment. Specifically, the fraction of inspired oxygen delivered to test subjects is varied to achieve a series of targeted steady state saturation periods over a range of arterial oxygen saturation of 70 - 100%.
While cerebral oximeters differ from pulse oximeters in terms of the what is being measured (brain tissue v arterial blood) the FDA have maintained the requirement to examine data from human volunteer studies in which the arterial oxygen saturation ranges from 70 - 100%. Two FDA-approved cerebral oximeters were validated in a similar manner.
The device controlling the inspired gas concentration is the RespirAct, which permits precise reduction in the arterial oxygen saturation while maintaining the arterial carbon dioxide level at 40 mmHg.
The study consists of 3 sequences:
- First sequence: reduction in arterial oxygen saturation in approximately 5% increments from 100 to 70% followed by return to room air and then a period of supplemental oxygen.
- Second sequence: reduction in arterial oxygen saturation in a single drop from 100 to 70% followed by return to room air and then a period of supplemental oxygen.
- Third sequence: reduction in arterial oxygen saturation in approximately 10% increments from 100 to 70% with alteration of carbon dioxide level from 35 to 45 mmHg at each interval followed by return to room air.
| Study Type : | Interventional (Clinical Trial) |
| Actual Enrollment : | 23 participants |
| Allocation: | N/A |
| Intervention Model: | Single Group Assignment |
| Masking: | None (Open Label) |
| Primary Purpose: | Diagnostic |
| Official Title: | Calibration and Validation of the 4 Wavelength Nonin Non-invasive Cerebral Oxygen Saturation Oximeter and Cerebral Sensor |
| Study Start Date : | September 2009 |
| Actual Primary Completion Date : | June 2010 |
| Actual Study Completion Date : | April 2012 |
| Arm | Intervention/treatment |
|---|---|
|
Experimental: Desaturation
Human volunteers undergo oxygen desaturation in order to determine the accuracy of the device over a clinical range of oxygen saturations 70 - 100%.
|
Device: Reduction of inspired oxygen
Reduction in blood oxygen saturation by sequential reduction in inspired gas composition. Steps are of 6 minutes duration with reduction in pulse oximeter oxygen saturation from 100 to 70%.
Other Name: Nonin Equanox cerebral oximeter |
- Accuracy of Sensor [ Time Frame: Data collected from individual participants over 1 hour timeframe. Data from cohort of subjects collected over 6 month period. ]A scatterplot is created with the forehead sensor saturation on the y-axis and the measured blood saturation on the x-axis. The line of identity is drawn representing the ideal points, meaning that the forehead sensor saturation is always the same as the blood saturation. The dispersion of the actual data points around this line of identity can be measured using a statistical calculation called Arithmetic Root Mean Square or ARMS. The smaller the ARMS the closer the data points lie around the line of identity, representing a more accurate sensor.
Choosing to participate in a study is an important personal decision. Talk with your doctor and family members or friends about deciding to join a study. To learn more about this study, you or your doctor may contact the study research staff using the contacts provided below. For general information, Learn About Clinical Studies.
| Ages Eligible for Study: | 21 Years to 35 Years (Adult) |
| Sexes Eligible for Study: | All |
| Accepts Healthy Volunteers: | Yes |
Inclusion Criteria:
- 21 to 35 years of age
- American Society of Anesthesiologists health assessment level 1
- Body Mass Index (BMI) 18 to 30
Exclusion Criteria:
- anemia
- hemoglobinopathy (e.g. sickle cell disease, thalassemia)
- positive pregnancy test (females)
- significant cardiac or pulmonary disease
- history of sleep apnea
- tobacco, drug or alcohol abuse
- difficult airway
- abnormal EKG / pulmonary function test / room air saturation
- intolerance to breathing mask apparatus
To learn more about this study, you or your doctor may contact the study research staff using the contact information provided by the sponsor.
Please refer to this study by its ClinicalTrials.gov identifier (NCT number): NCT01762722
| United States, North Carolina | |
| Duke University Health System | |
| Durham, North Carolina, United States, 27710 | |
| Principal Investigator: | David B MacLeod, FRCA | Duke Health |
| Responsible Party: | David MacLeod, Assistant Clinical professor, Duke University |
| ClinicalTrials.gov Identifier: | NCT01762722 |
| Other Study ID Numbers: |
Pro00018006 |
| First Posted: | January 8, 2013 Key Record Dates |
| Results First Posted: | April 16, 2013 |
| Last Update Posted: | April 16, 2013 |
| Last Verified: | March 2013 |
|
Spectroscopy, near infrared Oximetry Hypoxemia Tissue oxygenation |