Comparison Evaluation of Blood Glucose Meter Systems in a Clinical Setting

This study has been completed.
Sponsor:
Information provided by (Responsible Party):
Bayer HealthCare, Diabetes Care
ClinicalTrials.gov Identifier:
NCT01614613
First received: May 13, 2012
Last updated: September 9, 2013
Last verified: September 2013
  Purpose

The purpose of this study was to test the performance of an investigational Bayer Blood Glucose (BG) meter and five additional Blood Glucose Monitoring Systems (BGMSs). The goal was to obtain glucose samples from subjects with diabetes across the meter operating ranges. A capillary sample was collected from each subject at 3 different times during the visit to obtain natural capillary blood samples with a range of glucose concentrations. Some blood samples were glycolyzed to lower the glucose concentration levels and glucose solution was added to other samples to raise glucose concentration levels. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer). All lancings and testing were performed by study staff.


Condition Intervention
Diabetes
Device: Bayer G3 / Tatsu System
Device: Accu-Chek® Aviva Nano Meter/Accu-Chek® Aviva Test Strips
Device: Freestyle Lite® Meter and Test Strips with ZipwikTM tabs
Device: OneTouch® Ultra®2 / OneTouch® Ultra® Blue Test Strips
Device: One Touch® VerioTM Pro/One Touch® VerioTM Test Strips
Device: Truetrack® Meter/Truetrack® Test Strips

Study Type: Interventional
Study Design: Endpoint Classification: Efficacy Study
Intervention Model: Single Group Assignment
Masking: Open Label
Primary Purpose: Diagnostic
Official Title: Clinical Evaluation of Bayer G3 / Tatsu System and BGMs From Other Companies

Resource links provided by NLM:


Further study details as provided by Bayer HealthCare, Diabetes Care:

Primary Outcome Measures:
  • MAD Mean Absolute Value of the Difference Between Blood Glucose Meter (BGMS) Results and Corresponding YSI Blood Glucose (BG) Results in the Low BG Range(<70 mg/dL) [ Time Frame: 6 hours ] [ Designated as safety issue: No ]
    Using samples with BG < 70 mg/dL, the Mean Absolute Value of the Differences Between BGM system readings and the YSI laboratory reference values were compared. MAD was calculated from the sum of all /(BG meter)-(BG reference)/ assessments, divided by the number of assessments. Each evaluable sample was tested on all 6 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MAD value indicates smaller difference between meter value and the reference value. Higher MAD value indicates larger difference between meter value and the reference value.


Secondary Outcome Measures:
  • MAD Mean Absolute Value of the Difference Between Blood Glucose Meter (BGMS) Results and Corresponding YSI Blood Glucose (BG) Results in the BG Range 70 to 180 mg/dL [ Time Frame: 6 hours ] [ Designated as safety issue: No ]
    Using samples with BG 70 mg/dL to 180 mg/dL, the Mean Absolute Value of the Differences Between BGM system readings and the YSI laboratory reference values were compared. MAD was calculated from the sum of all/(BG meter)-(BG reference)/ assessments, divided by the number of assessments. Each evaluable sample was tested on all 6 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MAD value indicates smaller difference between meter value and the reference value. Higher MAD value indicates larger difference between meter value and the reference value.

  • MAD Mean Absolute Value of the Difference Between Blood Glucose Meter (BGMS) Results and Corresponding YSI Blood Glucose (BG) Results in the BG Range >180 mg/dL [ Time Frame: 6 hours ] [ Designated as safety issue: No ]
    Using samples with BG >180 mg/dL, the Mean Absolute Value of the Differences Between BGM system readings and the YSI laboratory reference values was compared. MAD was calculated from the sum of all /(BG meter)-(BG reference)/ assessments, divided by the number of assessments. Each evaluable sample was tested on all 6 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MAD value indicates smaller difference between meter value and the reference value. Higher MAD value indicates larger difference between meter value and the reference value.

  • Standard Deviations of BGMS Differences (Between BGM Meter Readings and the YSI Laboratory Reference Values Across the BG Range of All Evaluable Samples 21 mg/dL to 496 mg/dL) [ Time Frame: 6 hours ] [ Designated as safety issue: No ]
    Variability of each meter system was determined by calculating the Standard Deviation derived from each meter's differences between Blood Glucose Meter (BGMS)results and corresponding YSI Blood Glucose (BG) results.

  • Number of Subject Responses 'Strongly Agree' 'Agree' or 'Neutral' With Questionnaire Statements [ Time Frame: 1 hour ] [ Designated as safety issue: No ]
    Subjects responded to statements about meter accuracy and diabetes management. Subject responses: 'Strongly Agree' 'Agree' 'Neutral' 'Disagree' or 'Strongly Disagree'or No Response. □1ACCURACY HELPS 1A-with my ability to talk with my HCP. 1B-my satisfaction with my self monitoring of diabetes. 1C-with my ability to manage my diabetes. 1D-prevent low BG. 1E-understand how food/exercise affects low BG. □2 I WOULD USE ONLY the meter and strips my insurance company pays for, even if a more accurate meter was available. □3 I USE MY CURRENT METER BECAUSE 3A-my HCP gave it to me. 3B-my insurance company covers the strips. 3C-I think it is the most accurate meter. □4 I WOULD SWITCH meters for a more accurate meter. □5BEING ABLE TO APPLY MORE BLOOD IS IMPORTANT 5A-as I have wasted test strips by not having enough blood to fill the strip 5B-as this would save test strips 5C-I would switch to a meter with this feature


Enrollment: 146
Study Start Date: May 2012
Study Completion Date: June 2012
Primary Completion Date: June 2012 (Final data collection date for primary outcome measure)
Arms Assigned Interventions
Experimental: Intended BGM Users
Subjects were assigned to subgroups according to their initial glucose levels. SUBGROUP 1 goal: safely decrease subject glucose levels during the visit. SUBGROUP 2 goal: safely raise subject glucose levels. Also, blood samples were modified to obtain glucose concentrations across the needed glucose ranges while maintaining subject safety. Staff tested the blood samples using the 6 Blood Glucose Monitoring Systems. Bayer G3/Tatsu System;Accu-Chek® Aviva Nano Meter/Accu-Chek® Aviva Test Strips; Freestyle Lite® Meter and Test Strips with ZipwikTM tabs;OneTouch® Ultra®2 / OneTouch® Ultra® Blue Test Strips;One Touch® VerioTM Pro/One Touch® VerioTM Test Strips;Truetrack® Meter/Truetrack® Test Strips
Device: Bayer G3 / Tatsu System
Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
Device: Accu-Chek® Aviva Nano Meter/Accu-Chek® Aviva Test Strips
Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
Device: Freestyle Lite® Meter and Test Strips with ZipwikTM tabs
Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
Device: OneTouch® Ultra®2 / OneTouch® Ultra® Blue Test Strips
Study staff performed Blood Glucose (BG) tested with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
Device: One Touch® VerioTM Pro/One Touch® VerioTM Test Strips
Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
Device: Truetrack® Meter/Truetrack® Test Strips
Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

  Eligibility

Ages Eligible for Study:   18 Years and older
Genders Eligible for Study:   Both
Accepts Healthy Volunteers:   No
Criteria

Inclusion Criteria:

  • Diagnosis of type 1 or type 2 diabetes mellitus
  • 18 years of age or older
  • Willing to follow all study procedures

Exclusion Criteria:

  • Received an investigational drug or device within the last 30 days
  • Current use or recent exposure to any prescription medication within 24 hours prior to screening that in the opinion of the investigator could have an influence on the subject's ability to participate in the study or on the performance on the test device
  • Positive pregnancy test obtained at screening (test performed for females only, except if surgically sterile or three years past menopause)
  • Hematocrit <32% or >55%
  • Inadequate veins (in the opinion of the investigator or designated study staff) for venous blood withdrawal or intravenous catheter insertion
  • Symptomatic coronary artery disease with a history of angina, or history of a myocardial infarction or coronary intervention (e.g. percutaneous transluminal coronary angioplasty [PTCA], stent placement) or coronary artery bypass graft (CABG) within the past 6 months
  • Cerebrovascular accident within the past 6 months
  • Diagnosis of the following diabetic autonomic neuropathies: symptomatic orthostatic hypotension, heart rate anomalies, gastroparesis
  • Being treated for malignancy, except basal cell or squamous cell skin cancers
  • Major surgical operation within 30 days prior to screening
  • Active alcohol abuse or substance abuse (as judged by the principal investigator)
  • Severe mental illness, mental incapacity, unwillingness or language barriers precluding adequate understanding or cooperation (as judged by principal investigator)
  • Other medical conditions that would pose safety concerns, interfere with study conduct or seriously compromise study integrity, as judged by principal investigator (reason for exclusion will be clearly documented by investigator or designee)
  • Working for a competitive medical device company, or having an immediate family member or someone who is not a family member but is living within the household of someone who works for such a company
  Contacts and Locations
Choosing to participate in a study is an important personal decision. Talk with your doctor and family members or friends about deciding to join a study. To learn more about this study, you or your doctor may contact the study research staff using the Contacts provided below. For general information, see Learn About Clinical Studies.

Please refer to this study by its ClinicalTrials.gov identifier: NCT01614613

Locations
United States, Washington
Rainier Clinical Research Center Inc
Renton, Washington, United States, 98057
Sponsors and Collaborators
Bayer HealthCare, Diabetes Care
Investigators
Principal Investigator: Leslie J Klaff, MD Rainier Clinical Research Center Inc
  More Information

No publications provided

Responsible Party: Bayer HealthCare, Diabetes Care
ClinicalTrials.gov Identifier: NCT01614613     History of Changes
Other Study ID Numbers: CTD-2010-009-02
Study First Received: May 13, 2012
Results First Received: June 30, 2013
Last Updated: September 9, 2013
Health Authority: United States: Institutional Review Board