Performance of a New Glucose Meter System
This study has been completed.
Sponsor:
Bayer HealthCare, Diabetes Care
Information provided by (Responsible Party):
Bayer HealthCare, Diabetes Care
ClinicalTrials.gov Identifier:
NCT01264016
First received: December 18, 2010
Last updated: April 4, 2012
Last verified: April 2012
- Full Text View
- Tabular View
- Study Results
- Disclaimer
- How to Read a Study Record
Purpose
The purpose of this study is to evaluate the performance and acceptability of a new blood glucose monitoring system, which includes meter and sensor strip.
| Condition | Intervention |
|---|---|
|
Diabetes |
Device: Investigational Blood Glucose Monitoring System |
| Study Type: | Interventional |
| Study Design: | Endpoint Classification: Efficacy Study Intervention Model: Single Group Assignment Masking: Open Label Primary Purpose: Diagnostic |
| Official Title: | Performance of the TATSU/Tradewind Blood Glucose Monitoring System |
Resource links provided by NLM:
Further study details as provided by Bayer HealthCare, Diabetes Care:
Primary Outcome Measures:
- Percent of Capillary Blood Glucose (BG) Results Within +/- 5 to 20mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method [ Time Frame: 2 hours ] [ Designated as safety issue: No ]Subjects with diabetes used an investigational blood glucose monitoring system (BGMS) with subject capillary blood. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 20 mg/dL (for reference BG results <75mg/dL) or +/- 5 to 20% (for reference BG results >=75mg/dL) of the reference method results.
- Percent of Venous Blood Glucose (BG) Results Within +/- 5 to 20 mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method [ Time Frame: 2 hours ] [ Designated as safety issue: No ]Healthcare professionals (study staff) used an investigational blood glucose monitoring system (BGMS) with subject venous blood. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 20 mg/dL (for reference BG results <75mg/dL) or +/- 5% to 20% (for reference BG results >=75mg/dL) of the reference method results.
Secondary Outcome Measures:
- Number of Subjects Rated as<=3 Performing Basic Meter Tasks (Labeling Comprehension) [ Time Frame: 2 hours ] [ Designated as safety issue: No ]Subjects reviewed the instructions for use (User Guide and Quick Reference Guide) to learn to use the system. Study staff then observed and rated the subjects (1 to 4) on their success at performing five tasks. Scale: 1.Success in performing tasks correctly without assistance. 2.Successful with additional review of User Guide. 3.Successful with additional review and study staff assist similar to review of a specific function during a Customer Service call. 4.Subject did not perform task correctly and study staff intervention was required.
| Enrollment: | 96 |
| Study Start Date: | November 2010 |
| Study Completion Date: | December 2010 |
| Primary Completion Date: | December 2010 (Final data collection date for primary outcome measure) |
| Arms | Assigned Interventions |
|---|---|
|
Experimental: Intended Users of the Monitoring System
Untrained subjects with diabetes use an investigational blood glucose monitoring system.
|
Device: Investigational Blood Glucose Monitoring System
Untrained subjects with diabetes performed Blood Glucose (BG) tests with subject's capillary fingerstick using the investigational Tradewind meter and investigational sensor. Study staff tested subject venous blood. All BG Results were compared to a reference laboratory glucose method. Subjects' success at performing basic tasks using only the User Guide were rated by study staff.
|
Eligibility| Ages Eligible for Study: | 18 Years and older |
| Genders Eligible for Study: | Both |
| Accepts Healthy Volunteers: | No |
Criteria
Inclusion Criteria:
- Type 1 or 2 diabetes
- Age 18 years or older
- Routinely performs blood glucose self testing at home, at least once per day
- Able to speak, read, and understand English.
- Is willing to complete all study procedures
Exclusion Criteria:
- Currently pregnant
- Hemophilia or any other bleeding disorder
- Taking prescription anticoagulants or has clotting problems that may prolong bleeding. Taking aspirin daily is not excluded
- Subject employee of competitive medical device company
- Cognitive disorder or other condition, which in the opinion of the investigator, would put the person at risk or seriously compromise the integrity of the study
Contacts and Locations
Please refer to this study by its ClinicalTrials.gov identifier: NCT01264016
Locations
| United States, California | |
| AMCR Institute | |
| Escondido, California, United States, 92026 | |
Sponsors and Collaborators
Bayer HealthCare, Diabetes Care
Investigators
| Principal Investigator: | Timothy Bailey, MD | AMCR Institute |
More Information
No publications provided
| Responsible Party: | Bayer HealthCare, Diabetes Care |
| ClinicalTrials.gov Identifier: | NCT01264016 History of Changes |
| Other Study ID Numbers: | CTD-2010-004-01 |
| Study First Received: | December 18, 2010 |
| Results First Received: | April 4, 2012 |
| Last Updated: | April 4, 2012 |
| Health Authority: | United States: Institutional Review Board |
ClinicalTrials.gov processed this record on May 23, 2013