RestoreSensor Study

This study has been completed.
Sponsor:
Information provided by (Responsible Party):
MedtronicNeuro
ClinicalTrials.gov Identifier:
NCT01106404
First received: April 2, 2010
Last updated: August 17, 2012
Last verified: August 2012

April 2, 2010
August 17, 2012
April 2010
January 2011   (final data collection date for primary outcome measure)
Percentage of Subjects With Improved Pain Relief and/or Convenience During the AdaptiveStim Programming Arm Relative to the Manual Programming Arm [ Time Frame: 16 weeks post-implant ] [ Designated as safety issue: No ]
After subjects experienced both AdaptiveStim and manual programming at 16 weeks post-implant, subjects were asked to compare pain relief and convenience when they had AdaptiveStim ON to AdaptiveStim OFF in two separate domains using two 5-point Likert scales. The outcome measure for the primary objective is the percentage of subjects who report improved pain relief with no loss of convenience or improved convenience with no loss of pain relief during the AdaptiveStim programming arm relative to the manual programming arm. These subjects were considered successful for the primary objective.
Not Provided
Complete list of historical versions of study NCT01106404 on ClinicalTrials.gov Archive Site
  • Percentage of Subjects With Worsened Pain Relief When Using AdaptiveStim Compared to Manual Programming [ Time Frame: 16 weeks post-implant ] [ Designated as safety issue: No ]
    The pain relief question was a 5-point Likert scale question comparing pain relief when using AdaptiveStim programming relative to manual programming after subjects finished both programming periods. The choices were much worse pain relief with AdaptiveStim, somewhat worse pain relief with AdaptiveStim, no difference in pain relief, somewhat better pain relief with AdaptiveStim, and much better pain relief with AdaptiveStim. Subjects who had worsening pain relief were defined as subjects who responded "much worse pain relief with AdaptiveStim" or "somewhat worse pain relief with AdaptiveStim".
  • Manual Adjustments Presented as Button Presses [ Time Frame: Baseline, 10 weeks and 16 weeks post-implant ] [ Designated as safety issue: No ]
    The number of individual adjustments, ie, button presses, were recorded automatically in the patient programmer, which was specifically designed for this study. The number of button presses per day for manual patient programmer adjustments during the manual programming arm and the AdaptiveStim programming arm of the study were compared.
  • NPRS Scores From Baseline to Follow-up Visits at 10 Weeks and 16 Weeks Post-implant [ Time Frame: Baseline, 10 weeks and 16 weeks post-implant ] [ Designated as safety issue: No ]
    The 11-point (ie, 0-10) numeric rating scale of pain intensity (NPRS) was used to assess the overall pain at baseline and follow-up visits. In the pain diary, subjects were presented a numeric scale with numbers from 0 to 10, with 0 meaning "No pain" and 10 meaning "Pain as bad as you can imagine," accompanied by the instructions "Please rate your pain by indicating the number that best describes your pain on average in the last 24 hours." The subjects were asked to complete a diary for 7 consecutive days before implant, 10 weeks, and 16 weeks post-implant visits.
Not Provided
Not Provided
Not Provided
 
RestoreSensor Study
RestoreSensor Study

The purpose of the study is to gather clinical information on subjects' experiences with and without the use of the new feature of the RestoreSensor implantable neurostimulator (INS).

This study is a multi-center, prospective, open-label, randomized, crossover study that is designed to gather clinical information on subjects' experiences with the use of the new feature of the RestoreSensor implantable neurostimulator. Current spinal cord stimulators for chronic intractable pain stimulate targeted nerves along the spine using electrical impulses. The stimulation interferes with the transmission of pain signals to the brain replacing painful sensations with a tingling sensation called paresthesia. Variation in the intensity of neurostimulation with body position is a challenge for some patients implanted with conventional spinal cord stimulation systems because positional changes may result in overstimulation or understimulation. Patients need to manually adjust their stimulation using the patient programmer to maintain their comfort level during position changes. The AdaptiveStim feature of the RestoreSensor neurostimulator was developed to address this challenge by improving pain relief and/or convenience relative to manual programming adjustments.

Interventional
Not Provided
Allocation: Randomized
Endpoint Classification: Efficacy Study
Intervention Model: Crossover Assignment
Masking: Open Label
Primary Purpose: Treatment
Chronic Pain
Device: RestoreSensor Neurostimulation System
Adjustments to stimulation settings are automatic when the AdaptiveStim feature of the device is turned "ON" and manual when the AdaptiveStim feature of the device is turned "OFF".
  • 6-week AdaptiveStim followed by 6-week manual programming
    Intervention: Device: RestoreSensor Neurostimulation System
  • 6-week manual followed by 6-week AdaptiveStim programming
    Intervention: Device: RestoreSensor Neurostimulation System
Schultz DM, Webster L, Kosek P, Dar U, Tan Y, Sun M. Sensor-driven position-adaptive spinal cord stimulation for chronic pain. Pain Physician. 2012 Jan-Feb;15(1):1-12.

*   Includes publications given by the data provider as well as publications identified by ClinicalTrials.gov Identifier (NCT Number) in Medline.
 
Completed
79
January 2011
January 2011   (final data collection date for primary outcome measure)

Inclusion Criteria:

  • Meets the indications as an appropriate surgical candidate (after successful SCS screening trial) for implant of a neurostimulation system for spinal cord stimulation (SCS) treatment of trunk and/or limb pain, per labeled indications
  • 18 years of age or older
  • Willing and able to attend visits and comply with the study protocol
  • Capable of using the patient programmer and recharging the neurostimulator (INS), able to read and answer questionnaires in English, without assistance of a caregiver
  • Males and non-pregnant females

Exclusion Criteria:

  • Has had a prior implantable SCS neurostimulation system
  • Currently enrolled, or plans to enroll in another investigational device or drug trial during the study
  • Has unresolved legal issues related to their pain condition for which they would be receiving neurostimulation
  • Requires cervical placement of leads.
Both
18 Years and older
No
Contact information is only displayed when the study is recruiting subjects
United States
 
NCT01106404
1651
No
MedtronicNeuro
MedtronicNeuro
Not Provided
Not Provided
MedtronicNeuro
August 2012

ICMJE     Data element required by the International Committee of Medical Journal Editors and the World Health Organization ICTRP