The EVOLVE II Clinical Trial To Assess the SYNERGY Stent System for the Treatment of Atherosclerotic Lesion(s)
- Full Text View
- Tabular View
- No Study Results Posted
- Disclaimer
- How to Read a Study Record
Purpose
The purpose of this study is to assess the safety and effectiveness of the SYNERGY Everolimus-Eluting Platinum Chromium Coronary Stent System for the treatment of subjects with atherosclerotic lesion(s) ≤ 34 mm in length (by visual estimate) in native coronary arteries ≥2.25 mm to ≤4.0 mm in diameter (by visual estimate).
| Condition | Intervention | Phase |
|---|---|---|
|
Coronary Artery Disease |
Device: PROMUS Element Plus Device: SYNERGY |
Phase 3 |
| Study Type: | Interventional |
| Study Design: | Allocation: Randomized Endpoint Classification: Safety/Efficacy Study Intervention Model: Parallel Assignment Masking: Single Blind (Subject) Primary Purpose: Treatment |
| Official Title: | EVOLVE II: A Prospective Multicenter Trial to Assess the Safety and Effectiveness of the SYNERGY Everolimus-Eluting Platinum Chromium Coronary Stent System (SYNERGY Stent System) for the Treatment of Atherosclerotic Lesion(s) |
- Target Lesion Failure (TLF) Rate [ Time Frame: 12 months ] [ Designated as safety issue: Yes ]TLF is defined as any ischemia-driven revascularization of the target lesion, myocardial infarction (Q-wave and non-Q-wave) related to the target vessel, or cardiac death.
- Target Lesion Revascularization (TLR) Rate [ Time Frame: 30 days, 6 months, 12 months, 18 months, 2-5 years ] [ Designated as safety issue: No ]
- Target Lesion Failure (TLF) Rate [ Time Frame: 30 days, 6 months, 12 months, 18 months, 2-5 years ] [ Designated as safety issue: Yes ]
- Target Vessel Revascularization (TVR) Rate [ Time Frame: 30 days, 6 months, 12 months, 18 months, 2-5 years ] [ Designated as safety issue: No ]
- All Revascularization Rate [ Time Frame: 30 days, 6 months, 12 months, 18 months, 2-5 years ] [ Designated as safety issue: No ]
- Target Vessel Failure (TVF) Rate [ Time Frame: 30 days, 6 months, 12 months, 18 months, 2-5 years ] [ Designated as safety issue: Yes ]
- Myocardial Infarction (MI, Q-wave and non-Q-wave) Rate [ Time Frame: 30 days, 6 months, 12 months, 18 months, 2-5 years ] [ Designated as safety issue: Yes ]
- Cardiac Death Rate [ Time Frame: 30 days, 6 months, 12 months, 18 months, 2-5 years ] [ Designated as safety issue: Yes ]
- Non-Cardiac Death Rate [ Time Frame: 30 days, 6 months, 12 months, 18 months, 2-5 years ] [ Designated as safety issue: Yes ]
- All Death Rate [ Time Frame: 30 days, 6 months, 12 months, 18 months, 2-5 years ] [ Designated as safety issue: Yes ]
- Cardiac Death or Myocardial Infarction (MI) Rate [ Time Frame: 30 days, 6 months, 12 months, 18 months, 2-5 years ] [ Designated as safety issue: Yes ]
- All Death or Myocardial Infarction (MI) Rate [ Time Frame: 30 days, 6 months, 12 months, 18 months, 2-5 years ] [ Designated as safety issue: Yes ]
- All Death, Myocardial Infarction (MI) or Target Vessel Revascularization (TVR) Rate [ Time Frame: 30 days, 6 months, 12 months, 18 months, 2-5 years ] [ Designated as safety issue: Yes ]
- Stent Thrombosis Rate by Academic Research Consortium (ARC) Definition [ Time Frame: 30 days, 6 months, 12 months, 18 months, 2-5 years ] [ Designated as safety issue: Yes ]
- Stroke Rate (ischemic and hemorrhagic) [ Time Frame: 30 days, 6 months, 12 months, 18 months, 2-5 years ] [ Designated as safety issue: Yes ]
- Technical Success Rate [ Time Frame: Day 1 (periprocedure) ] [ Designated as safety issue: No ]
- Clinical Procedural Success Rate [ Time Frame: Day 1 (periprocedure) ] [ Designated as safety issue: Yes ]
| Estimated Enrollment: | 1684 |
| Study Start Date: | November 2012 |
| Estimated Study Completion Date: | June 2018 |
| Estimated Primary Completion Date: | June 2014 (Final data collection date for primary outcome measure) |
| Arms | Assigned Interventions |
|---|---|
|
Active Comparator: Promus Element Plus
PROMUS Element Plus is a device/drug combination product composed of two components, a device (coronary stent system including a platinum chromium stent platform) and a drug product (a formulation of everolimus contained in a polymer coating).
|
Device: PROMUS Element Plus |
|
Experimental: SYNERGY
SYNERGY is a device/drug combination product composed of two components, a device (coronary stent system including a platinum chromium stent platform) and a drug product (a formulation of everolimus contained in a bioabsorbable polymer coating).
|
Device: SYNERGY |
Eligibility| Ages Eligible for Study: | 18 Years and older |
| Genders Eligible for Study: | Both |
| Accepts Healthy Volunteers: | No |
Inclusion Criteria:
- Subject must be at least 18 years of age
- Subject (or legal guardian) understands the trial requirements and the treatment procedures and provides written informed consent before any trial-specific tests or procedures are performed
- For subjects less than 20 years of age enrolled at a Japanese site, the subject and the subject's legal representative must provide written informed consent before any study-specific tests or procedures are performed
- Subject is eligible for percutaneous coronary intervention (PCI)
- Subject has symptomatic coronary artery disease with objective evidence of ischemia or silent ischemia
- Subject is an acceptable candidate for coronary artery bypass grafting (CABG)
- Subject is willing to comply with all protocol-required follow-up evaluation
Angiographic Inclusion Criteria (visual estimate):
- Target lesion(s) must be located in a native coronary artery with a visually estimated reference vessel diameter (RVD) ≥2.25 mm and ≤4.0 mm
- Target lesion(s) length must be ≤34 mm (by visual estimate)
- Target lesion(s) must have visually estimated stenosis ≥50% and <100% with thrombolysis in Myocardial Infarction (TIMI) flow >1 and one of the following: stenosis ≥70%, abnormal fractional flow reserve (FFR), abnormal stress or imaging stress test, or elevated biomarkers prior to the procedure
- Coronary anatomy is likely to allow delivery of a study device to the target lesions(s)
- The first lesion treated must be successfully predilated/pretreated
Exclusion Criteria:
- Subject has clinical symptoms and/or electrocardiogram (ECG) changes consistent with acute ST elevation MI (STEMI)
- Subject has cardiogenic shock, hemodynamic instability requiring inotropic or mechanical circulatory support, intractable ventricular arrhythmias, or ongoing intractable angina
- Subject has received an organ transplant or is on a waiting list for an organ transplant
- Subject is receiving or scheduled to receive chemotherapy within 30 days before or after the index procedure
- Planned PCI (including staged procedures) or CABG after the index procedure
Subject previously treated at any time with intravascular brachytherapy
_ Subject has a known allergy to contrast (that cannot be adequately premedicated) and/or the trial stent system or protocol-required concomitant medications (e.g., platinum, platinum-chromium alloy, stainless steel, everolimus or structurally related compounds, polymer or individual components, all P2Y12 inhibitors, or aspirin)
Subject has one of the following (as assessed prior to the index procedure):
- Other serious medical illness (e.g., cancer, congestive heart failure) with estimated life expectancy of less than 24 months
- Current problems with substance abuse (e.g., alcohol, cocaine, heroin, etc.)
- Planned procedure that may cause non-compliance with the protocol or confound data interpretation
- Subject is receiving chronic (≥72 hours) anticoagulation therapy (i.e., heparin, coumadin) for indications other than acute coronary syndrome
- Subject has a platelet count <100,000 cells/mm3 or >700,000 cells/mm3
- Subject has a white blood cell (WBC) count < 3,000 cells/mm3
- Subject has documented or suspected liver disease, including laboratory evidence of hepatitis
- Subject is on dialysis or has baseline serum creatinine level >2.0 mg/dL (177µmol/L)
- Subject has a history of bleeding diathesis or coagulopathy or will refuse blood transfusions
- Subject has had a history of cerebrovascular accident (CVA) or transient ischemic attack (TIA) within the past 6 months
- Subject has an active peptic ulcer or active gastrointestinal (GI) bleeding
- Subject has severe symptomatic heart failure (i.e., NYHA class IV)
- Subject is participating in another investigational drug or device clinical trial that has not reached its primary endpoint
- Subject intends to participate in another investigational drug or device clinical trial within 12 months after the index procedure
- Subject with known intention to procreate within 12 months after the index procedure (women of child-bearing potential who are sexually active must agree to use a reliable method of contraception from the time of screening through 12 months after the index procedure)
- Subject is a woman who is pregnant or nursing (a pregnancy test must be performed within 7 days prior to the index procedure in women of child-bearing potential)
Angiographic Exclusion Criteria (visual estimate):
- Planned treatment of more than 3 lesions
- Planned treatment of lesions in more than 2 major epicardial vessels
- Planned treatment of a single lesion with more than 1 stent
- Subject has 2 target lesions in the same vessel that are separated by less than 15 mm (by visual estimate)
- Target lesion(s) is located in the left main
- Target lesion(s) is located within 3 mm of the origin of the left anterior descending (LAD) coronary artery or left circumflex (LCx) coronary artery by visual estimate
- Target lesion(s) is located within a saphenous vein graft or an arterial graft
- Target lesion(s) will be accessed via a saphenous vein graft or arterial graft
- Target lesion(s) with a TIMI flow 0 (total occlusion) or TIMI flow 1 prior to guide wire crossing
- Target lesion(s) treated during the index procedure that involves a complex bifurcation (e.g., bifurcation lesion requiring treatment with more than 1 stent)
- Target lesion(s) is restenotic from a previous stent implantation or study stent would overlap with a previous stent
- Subject has unprotected left main coronary artery disease (>50% diameter stenosis)
- Subject has been treated with any type of PCI (i.e., balloon angioplasty, stent, cutting balloon atherectomy) within 24 hours prior to the index procedure
- Thrombus, or possible thrombus, present in the target vessel (by visual estimate)
Contacts and Locations| Contact: Alison J Osattin, MPH | 508.683.6574 | Alison.Osattin@bsci.com |
Show 60 Study Locations| Study Director: | Peter M Maurer, MPH | Boston Scientific Corporation |
More Information
No publications provided
| Responsible Party: | Boston Scientific Corporation |
| ClinicalTrials.gov Identifier: | NCT01665053 History of Changes |
| Other Study ID Numbers: | S2067, G120123 |
| Study First Received: | August 7, 2012 |
| Last Updated: | April 4, 2013 |
| Health Authority: | United States: Food and Drug Administration |
Keywords provided by Boston Scientific Corporation:
|
Drug-Eluting Stents |
Additional relevant MeSH terms:
|
Coronary Artery Disease Myocardial Ischemia Coronary Disease Heart Diseases Cardiovascular Diseases Arteriosclerosis Arterial Occlusive Diseases |
Vascular Diseases Everolimus Immunosuppressive Agents Immunologic Factors Physiological Effects of Drugs Pharmacologic Actions |
ClinicalTrials.gov processed this record on June 17, 2013